11 Error messages
99
11.4 Measured value recording
EEPROM access
Internal diagnos-
tics
An error occurred during EEPROM reading/writing.
Restart device / return device.
USB communication
Internal diagnos-
tics
Error during communication via USB. Restart device / return device.
RS485 communication
Internal diagnos-
tics
Error during RS485 communication. Restart device / return device.
Display Cause/remedy
<<<<< Underrange / check sensor configuration, check measuring chain
>>>>>
Overrange / check sensor configuration, check measuring chain
- - - - - Invalid value / restart device, otherwise return device
- - 4 - -
Division by zero / restart device, otherwise return device
+++++
Error relating to terminal temperature or compensation signal / restart device, otherwise return device
<-<-<
Probe short circuit / check sensor configuration, check line for short circuit
>->-> Probe break / check sensor configuration, check line for short circuit
-1-0-
No data received from channel / restart device, otherwise return device
*****
Value cannot be displayed, overrange / restart device, otherwise return device
Display Origin Cause/remedy