4 Changing the Default Settings
112 Keysight U1253B User’s and Service Guide
2
T-TYPE K Thermocouple type.
Available options: K-type or J-type
T-UNIT °C Temperature unit.
Available options:
°C/°F: Dual display, °C in primary display, °F in secondary.
°C: Single display, in °C only.
°F/°C: Dual display, °F in primary display, °C in secondary.
°F: Single display, in °F only.
Press to swap between °C and °F.
mA-SCALE 4 mA to 20 mA Percentage scale for mA.
Available options: 4 – 20 mA, 0 – 20 mA, or OFF.
Select OFF to disable this function for normal operation.
CONTINUITY SINGLE Audible continuity.
Available options: SINGLE, OFF, or TONE.
MIN-Hz 0.5 Hz Minimum measurement frequency.
Available options: 0.5 Hz, 1 Hz, 2 Hz, or 5 Hz.
3
BEEP 2400 Beep frequency.
Available options: 4800 Hz, 2400 Hz, 1200 Hz, 600 Hz, or OFF.
To disable this function, select OFF.
APO 10 M Automatic power off.
To enable this function, select a value within the range of 1 minute to
99 minutes.
To disable this function, set all digits to zero (“OFF” will be indicated).
BACKLIT HIGH Default power-on backlight brightness level. Available options: HIGH, MEDIUM,
or LOW.
MELODY FACTORY Power-on melody. Available options: FACTORY, USER, or OFF.
GREETING FACTORY Power-on greeting. Available options: FACTORY, USER, or OFF.
Table 4-1 Default factory settings and available setting options for each feature (continued)
Menu Feature
Default factory
setting
Available setting options

Table of Contents