84 Laborie alpHaONE User & Service Manual LBL-01791 [B]
Appendix D: Electromagnetic Compatibility
(EMC)
The pH Capsule System needs special precautions with regard to Electromagnetic Compatibility (EMC), and it needs
to be installed and placed into service according to the EMC information provided in this manual.
TheusershouldassurethatthemodelJSPH-3isusedintheelectromagneticenvironmentspeciedinthetablesthat
follow.
TableD-1: Guidance and Manufactures Declaration—Electromagnetic Emissions
EmissionTest Compliance ElectromagneticEnvironmentGuidance
RF emissions
CISPR 11
Group 1 ThemodelJSPH-3usesRFenergyonlyforitsinternalfunction.
Therefore,itsRFemissionsareverylowandarenotlikelyto
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B ThemodelJSPH-3issuitableforusedindomesticestablishment
and in establishment directly connected to a low voltage power
supplynetworkwhichsuppliesbuildingsusedfordomestic
purposes.
Harmonic emissions
IEC61000-3-2
Not applicable
Voltageuctuations/
ickeremissions
IEC61000-3-3
Not applicable
TableD-2: Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
Immunity Test IEC60601TestLevel Compliance Level ElectromagneticEnvironmentGuidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floors should be wood, concrete, or ceramic tile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kVforinput/outputlines
±1kVforACandDC
powerlines
±1kVforinput/outputlines Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1kVlinetoline
±2kVlinetoearth
Not applicable Mains power quality should be that of a typical
commercial or hospital environment.
Continued on next page