84 Laborie alpHaONE User & Service Manual LBL-01791 [B]
Appendix D: Electromagnetic Compatibility
(EMC)
The pH Capsule System needs special precautions with regard to Electromagnetic Compatibility (EMC), and it needs
to be installed and placed into service according to the EMC information provided in this manual.
TheusershouldassurethatthemodelJSPH-3isusedintheelectromagneticenvironmentspeciedinthetablesthat
follow.
TableD-1: Guidance and Manufacture’s Declaration—Electromagnetic Emissions
EmissionTest Compliance ElectromagneticEnvironmentGuidance
RF emissions
CISPR 11
Group 1 ThemodelJSPH-3usesRFenergyonlyforitsinternalfunction.
Therefore,itsRFemissionsareverylowandarenotlikelyto
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class B ThemodelJSPH-3issuitableforusedindomesticestablishment
and in establishment directly connected to a low voltage power
supplynetworkwhichsuppliesbuildingsusedfordomestic
purposes.
Harmonic emissions
IEC61000-3-2
Not applicable
Voltageuctuations/
ickeremissions
IEC61000-3-3
Not applicable
TableD-2: Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
Immunity Test IEC60601TestLevel Compliance Level ElectromagneticEnvironmentGuidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6kVcontact
±8kVair
±6kVcontact
±8kVair
Floors should be wood, concrete, or ceramic tile.
Ifoorsarecoveredwithsyntheticmaterial,the
relativehumidityshouldbeatleast30%.
Electrical fast
transient/burst
IEC 61000-4-4
±2kVforinput/outputlines
±1kVforACandDC
powerlines
±1kVforinput/outputlines Mains power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
±1kVlinetoline
±2kVlinetoearth
Not applicable Mains power quality should be that of a typical
commercial or hospital environment.
Continued on next page