LBL-01791 [B] Laborie alpHaONE User & Service Manual 89
Appendix F: Software Messages
Inthisappendix,youcanndanoverviewofthesoftwaremessagesaswellasmessagesthatcandisplayonthe
alpHaONE recorder, along with possible action(s) to resolve.
TableF-1: Troubleshooting the alpHaONE pH Capsule
Fault Cause Possible Actions to Resolve
Capsuleisnotblinkingafter
breakerremoval
Expired capsule Replace the capsule.
ActivationFailed Expired capsule Replace the capsule.
Low signal Move the capsule closer to the recorder.
Signal interference Makesuretherearenootheractivecapsules(includingother
brands) nearby.
Breakerwasremovedtooearly Placebreakerandthenremoveitandsearchagain.
Capsulexationfailed Grippinghookclosed Replace the capsule.
Capsule is not at the proper
position
Measuringthexingpositionaccordingtotheinstructions.
Improper operation When the capsule arrives at the designated position, do not
movethedeliverysystemuntilthesuctionprocessisnished.
Low pressure Keep the pressure at 0.08 MPa (600 mmHg), and maintain for 10
seconds.
The conveyer cannot be
takenoutafterthecapsuleis
axed
— Refer to AppendixB,“HandlingProcedurefortheDelivery
System” on page 82.
TableF-2: Troubleshooting the alpHaONE Data Recorder
Fault Cause Possible Actions to Resolve
Recorder does not turn on
after batteries are installed
Batteries failure Checkwhetherthebatteriesareinstalledcorrectly.
Low or failed batteries Replace the batteries with a new set.
Recorder failure Change the recorder.
Recorder shuts down
immediately after turning on
Low battery Replacethebatterieswithanewset;ifthisissuepersists,
contact the Laborie service department.
Low signal Move the capsule closer to the recorder.
Continued on next page