90 Laborie alpHaONE User & Service Manual LBL-01791 [B]
Fault Cause Possible Actions to Resolve
Unstable signal level Magnetic interference Keeptherecorderawayfromthemagneticeld.
The operational distance
between the capsule and the
recorder has been exceeded
Keep capsule and recorder within three meters of each other.
The capsule is already excreted
from the body
The examination is complete.
TableF-3: alpHaONE Recorder Display Messages
In the event of improper operation on the part of the user or an abnormal situation, the recorder emits a single beep and displays
the related message on its screen, as presented in this table.
Message Cause Possible Actions to Resolve
Battery low, please change battery Low-battery alarm Replace the batteries with a new set.
Memory chip error, please contact
manufacturer for support
Memory chip error Please contact the Laborie service
department.
Clockerror,pleasesetdateand
time
Date/time settings lost due to prolonged
storage
Reprogram the recorder for a new study to
reset data/time.
No capsule signal! The recorder does not receive a signal from
the capsule
Keeptherecorderawayfrommagneticelds.
Ensure the recorder is positioned close to the
patient during the study.
No capsule signal! Search again? Capsule signal is not received while
activating the capsule
Search capsule again.
Study completed Analertthatthestudyhadbeencompleted —
TableF-4: alpHaONE Software Messages
Message Cause Posible Actions to Resolve
Communication test failed, please try
again
N/A Contact your Laborie representative.
Continue with transferring? Continue transfer after determining invalid
patient number from alpHaONE
Actionrequiredfromoperator.
Illegal use of Laborie software on a
network
Applicationcannotbestartedfrom
networklocation
Start the application from the Laborie
Database module.
Internal error: Invalid patient number
from Database. Contact your dealer.
Invalid patient number from database Contact your Laborie representative.
Internal error: Invalid patient number
from alpHaONE. Contact your dealer!
Data will be transferred as a new
investigation for current patient in
database.
Invalid patient number from alpHaONE Contact your Laborie representative.
This color scheme is read-only Operator tries to change a system color
scheme
Modify a different color scheme.
Verify: Programming failed! Programming investigation protocol failed Contact your Laborie representative.