Recommendations
• If there is a risk that the system is overloaded and therefore shuts down thermally, LINAK recommends using quick release actuators. These will allow
functions to be lowered manually in case of a PJ2 malfunction due to misuse/abuse.
• If the customer has other essential performance than “no unintended movement”, he must consider this in his own risk analysis. LINAK disclaims
any liability.
• If the actuator or the control cable is removed from the control box, the cable lock must be applied. To ensure movement in this case, LINAK
recommends using quick release actuators in the application.
• To avoid cables from being damaged by pulling, LINAK recommends making safe cabling. If movement is an essential performance, LINAK
recommends applying quick release actuators, for example, to ensure movement.
• To avoid activation of thermal protection, do not exceed load specifications. If movement is an essential performance, LINAK recommends applying
quick release actuators, for example, to ensure movement.
• Sales must request a review of the products according to current cut-off limits.
• Push plugs fully into correct sockets and make sure that the plugs are completely inserted.
• Mount control box lid and close lid until locked in place.
Warnings
• Use EPR or ensure that the user takes care not to squeeze the mains cable.
• Always check correct assembly after mounting and service to ensure that the cable lock is mounted. (Connectors are usually removed during
cleaning)
• Always use approved chemicals with the housing as the plastic may show corrosion caused by some chemicals. As a result water may accumulate/
gather in housing.
• Take special precautions concerning 3
rd
party interfacing. Please contact LINAK for further information.
• Make a review of all product specifications before system set-up if the current cut-off limit is higher than the maximum allowed current cut-off for
the actuator.
• To avoid cable interruption and actuator defects make a proper cable installation and inspect regularly for wear and damage. Defective parts must
be replaced.
• After service inspection, the application must be tested for correct functionality before it is put into operation to avoid actuator plugs being mixed
during service. Operators must not be inside entrapment area.
• To avoid electrical failure or system disturbance inspect regularly for wear and damage. Defective parts must be replaced.
• Make a proper cable installation to avoid short-circuit cables for handset/controls. Regular inspection must be made for wear and damage. Defective
parts must be replaced.
• Do not mount the actuator with the spindle facing downwards to avoid that the actuator slips off the bracket with the mounted PJ2. The bracket can
come loose when exposed to shock or hard vibration, for instance when passing doorsteps. Regular inspection must be made to ensure proper
fixation of control box and bracket on actuator.
• PJ2 is ONLY to be used with PCP 2.0 control boxes.