Page 288 of 295
Recommendations
• The control buttons of redundant hand controls for lifting and lowering work as normal hand control buttons.
• The service counter is only active in COL50 when a sufficiently charged BAL50 is mounted.
• Only use original LINAK
®
mains cables to ensure a proper connection to the internal charger.
• When charging, LIFT50 will not be able to operate any actuators.
• Only use correct LINAK charger (CHL50 or integrated charger in LIFT50).
• To avoid cables from being damaged by pulling, LINAK recommends to make safe cabling.
• Push mains cable fully into correct sockets and make sure that the plugs are completely inserted.
Motor cable
Always use 6-wire cables. Please note that angled motor cable plugs are required for connection to the control box.
Hot Plugging
Removing or adding any OpenBus cables is not allowed when the control box is on power via mains supply or battery!
If necessary anyway, follow the below procedure:
1. Remove mains or battery and wait 5 sec.
2. Mount or dismount the required cables
If this procedure is NOT followed, it may result in a damaged OpenBus driver circuit.
The risk of a damaged circuit increases if the accessory has a high start current (in rush current).
Warnings
• Always check correct assembly after mounting and service to ensure that the cable lock is mounted. (Connectors are usually removed during cleaning)
• Always use approved chemicals with the housing as the plastic may show corrosion caused by some chemicals. As a result water may accumulate/
gather in housing.
• When using a control box with emergency stop, the stop button must be activated in cleaning situations to avoid unintended operation of the lift.
• Take special precautions concerning 3
rd
party interfacing. Please contact LINAK for further information.
• Make a review of all product specifications before system set-up if the current cut-off limit is higher than the maximum allowed actuator current cut-off.
• Make a proper cable installation and inspect regularly for wear and damage to avoid cable interruption and actuator defects.
Defective parts must be replaced.
• After service inspection, the application must be tested for correct functionality before it is put into operation to avoid actuator plugs being mixed
during service. Operators must not be inside entrapment area.
• Inspect regularly for wear and damage to avoid electrical failure or system disturbance. Defective parts must be replaced.
• Make a proper cable installation to avoid short-circuit cables for hand controls and other controls. Regular inspection must be made for wear and
damage. Defective parts must be replaced.
• To avoid injury, the battery should not be mounted in (all) transport situations. Use LINAK original packaging to store battery during transportation.
• Wireless: Be aware that a hand control can run an application that is out of sight.