User Manual 77
VMS4SE(Vibration Monitor)
REF NO: mVMA/om/101
Issue No: 01
Table 49 Holding Registers Parameters Part 4
Sr.
No
Parameter Description
Modbus
Address
Parameter
Type
Min
Value
Max
Value
Access
Remarks
1
Channel 1 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42201
42202
42203
Integer
-
-
R / W
Refer
Table 50
and Table
51 and
note.
2
Channel 2 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42204
42205
42206
Integer
-
-
R / W
3
Channel 3 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42207
42208
42209
Integer
-
-
R / W
4
Channel 4 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42210
42211
42212
Integer
-
-
R / W
5
Channel 5 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42213
42214
42215
Integer
-
-
R / W
6
Channel 6 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42216
42217
42218
Integer
-
-
R / W
7
Channel 7 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42219
42220
42221
Integer
-
-
R / W
8
Channel 8 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42222
42223
42224
Integer
-
-
R / W
9
Channel 9 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42225
42226
42227
Integer
-
-
R / W
10
Channel 10 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42228
42229
42230
Integer
-
-
R / W
11
Channel 11 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42231
42232
42233
Integer
-
-
R / W
12
Channel 12 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42234
42235
42236
Integer
-
-
R / W
13
Channel 13 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42237
42238
42239
Integer
-
-
R / W
14
Channel 14 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42240
42241
42242
Integer
-
-
R / W
15
Channel 15 ASCII code for
Digit -1 (Left Most Alpha Numeric
Digit) to Digit -6 (Right Most Alpha
Numeric Digit)
42243
42244
42245
Integer
-
-
R / W
16
Channel 16 ASCII code for
42246
Integer
-
-
R / W