TA80V User Manual
9 Ensure:
The end-expiratory pressure less than 3cmH
2
O
The ventilator displays the correct data
The bellows inflated and deflated during mechanical ventilation
10 Set the ventilator control and alarm limits to the appropriate clinical level
11 If the system is not used at the time, turn off the power switch
12 Make sure you have:
Airway maintenance
Manual ventilation
trachea cannula
Applicable anesthetic and
emergency medicines.
System preparation:
Turn off all of the vaporizer
Open the APL valve
Set the Auto/Manual switch to Manual
Set all flow control switches to a minimum
Ensure that the respiratory system is properly connected and remains intact
Warning
Ensure that the respiratory system is properly connected and remains intact.
Warning
Before connecting the device to the animal, flush the anesthesia machine with O2 at a flow rate of 5
L/min for at least one minute. This can remove unnecessary mixed gases and debris from the
system.
Warning
Anesthesia equipment must be connected to an exhaust gas system to discharge exhaust gas and
prevent harm to the health of operating room staff. This requirement needs to be adhered to whether
for testing or clinical applications.