Chapter 5 Installation and Connection
Attention
Our company strongly recommends: O2 monitoring should be used on this
equipment. Please refer to local standards for monitoring regulations.
Attention In accordance with the European standard EN 740 and the international standard
IEC 60601-2-13, this device shall use expiratory volume monitoring, O2 monitoring
(according to EN 12598 or ISO 7767).
Attention The European standard EN 740 and the international standard IEC 60601-2-13
also specify that when anesthetic vaporizers are used, anesthetic agent monitoring
should be carried out (in accordance with ISO 21647:2004).
Warning
Exhaled air can have an impact on the operating room environment. When
anesthetics remain untested for a long time, some unexpected dangers may arise.
The operator should handle the exhaled gas in a timely manner as required and
check other items. In this way, hazards and equipment failures should be reduced.
Warning Always ensure that the pipeline gas supply hose and breathing circuit components
are non-toxic and will not:
Cause an allergic reaction in animal patients.
Hazardous by-product produced in response to anesthetic gas or anesthetics.
Warning To prevent incorrect values or equipment failure, use the cables, hoses and piping
provided by us.
Warning
The drying absorber can be dangerous if there is an anesthetic. Appropriate
precautions should be taken to ensure that the sodium lime in the absorber is not
dry. After using the system, turn off all gas sources.
Warning To avoid false alarm caused to high strength electric field:
Place the electrosurgical lead away from the respiratory system and from the flow
and oxygen sensors.
Do not place the electrosurgical lead on any part of the anesthesia system.
Warning To protect animal patients, when using electrosurgical equipment, they shall:
Supervision ensures that all life support and monitoring equipment are operated
correctly.
In case the electrosurgical equipment cannot guarantee the safe use of the
ventilator, the standby artificial ventilation equipment should be readily available.
Do not use a conductive face mask or a hose.