Disinfection of pet patients with special infections or infections after use
Including open tuberculosis, lung abscess, Pseudomonas aeruginosa infection, tetanus, gas
gangrene, or infectious hepatitis. After use, all integrated parts of the bellows must be thoroughly
disinfected in two steps: preliminary treatment and thorough treatment.
1. Preliminary treatment: According to the principle of isolation treatment. Place all the integrated
parts of the bellows that have been used during the operation in the operating room. After the
operation, perform the following preliminary treatments: soak the integrated parts of the bellows with
1:1000 neo-germide or 1-5% cresol for 30 minutes.
2.Thorough treatment: After the above-mentioned preliminary treatment, the integrated parts of
the bellows are thoroughly disinfected:
a) Wash with soapy water, rinse with clean water repeatedly, and dry;
b) When conditions permit, it is best to fumigate the parts that are in direct contact with pet
patients with formaldehyde or ethylene oxide, or use soaking and disinfection respectively. For
example, parts used by pet patients with open tuberculosis should be soaked in 3% cresol for 30
minutes; pet patients with tetanus should be soaked with 0.2% potassium permanganate for 30
minutes; parts used by pet patients with gas gangrene, Soak the parts with 0.1% chlorhexidine for 30
minutes; soak the parts in pet patients with pulmonary purulent infection with 0.1% neocerin for 60
minutes; soak the parts in pet patients with Pseudomonas aeruginosa infection with 0.1% neocerin
for 2 hours;
c) After taking out the soaked objects, they need to be rinsed repeatedly with clean water and
dried for later use;
d) The parts that are not in direct contact with pet patients should be rubbed and rinsed
repeatedly with 1 to 3% phenol solution or soapy water and clean water. If necessary, irradiate with
ultraviolet light for 30 minutes.
6.5.5. Regular Maintenance
Warning
Do not perform any tests and repairs while it is being used on animal patients
to avoid endangering pet patients.
The following inspections should be carried out every 30 days to ensure that the parts that fail due to
use, daily cleaning, etc. are replaced in time.
Visual inspection
Disconnect the bellows integrated with the anesthesia machine.
Remove the bellows integration
Warning Do not separate the relief valve diaphragm and valve seat
Carefully inspect each component to determine if cracks, curling, dissolution, swelling and other
physical changes are present, replace if necessary, assemble, and then perform a leak test.