TA80V User Manual
8.3. Troubleshooting
8.3.1. Troubleshooting and Analysis of TA80V Anesthesia Machine
Leak in breathing circuit
The soda lime tank is not
installed tightly
Remove the soda lime
particles at the interface of
the soda lime tank
The threaded pipe is
damaged or the joint is loose
Replace with new pipe or
install again
Inhalation and exhalation
valve cover is not tightened
Manual/auto switch valve
malfunction
Please contact our agent or
us
Excessive pressure in the
airway when manual
ventilation
APL valve (pressure limiting
valve)
Regulated relief pressure is
too large
Adjust to proper relief
pressure
When the manual/auto
control switch is set to
manual, the airbag is
inflated; to auto, the bellow
inflated.
Manual/auto switch leakage
Please contact our agent or
us
There is serious leakage
in the respiratory system,
and the location of the
leakage is difficult to
determine at the moment
(in manual mode)
Improper air bag hose
connection.
Make sure the air bag hose
is connected to the air bag
port (under the APL valve).
8.3.2. Troubleshooting and Analysis of TA80V Ventilator
The power indicator light
not light up
Power cable not connected
Power cable damaged
The outlet to which the power
cable is connected has no
power.
The fuse is broken
Connect the power cable
Replace power cable
Use another power outlet
Replace fuse
A certain power socket is
out of power