Chapter 8 Alarm and Trouble-shooting
Warning
Personnel without experience in such equipment maintenance must never make
maintenance work.
8.1. About the Alarm
Attention If an alarm occurs, the animal safety should be protected first, and then fault
diagnosis or necessary maintenance procedures should be carried out.
Alarm information is displayed in the top area of the display
Figure 8-1 Alarm information Bar
High-priority alarms need to be handled immediately.
priority
The alarm volume
Alarm silent
Alarm information area
prompts
alarm indicator
High
5 alarm sounds, 2 of which
are rapid, and the alarm
period is 9 seconds.
120
Seconds
The red-bottom cycle
is shown.
Red, flashing.
Midle
3 tones, alarm period is 6
seconds.
120
Seconds
The yellow-bottom
circulation is shown.
Yellow, flashing.
Low
2 tones, with an alarm
period of 27 seconds.
120
Seconds
True color and low
cycle display.
Attention
When muted, the alarm bell changes to and the alarm sound disappears. After
2 minutes, the alarm bell returns to its original state; if the conditions for the alarm
are not processed, the alarm sound will continue to sound..
8.2. Alarm Information Table
Attention Protect animal patients during use; repair the fault after use.
Attention This table does not include the operating instructions.
Display
information
Alarm
level
Causes
Measures / precautions
Repair
AC power
failure
Middle
The cable not properly
connected
Power power outage
blown fuse
Check whether the cable
connection correct
Check if the power
supply interrupted
Check if the fuse
broken
If the fuse is
broken,
replace
immediately.