TA80V User Manual
Chapter 8 Alarm and Trouble-shooting
Warning
Personnel without experience in such equipment maintenance must never make
maintenance work.
8.1. About the Alarm
Attention If an alarm occurs, the animal safety should be protected first, and then fault
diagnosis or necessary maintenance procedures should be carried out.
Alarm information is displayed in the top area of the display
Figure 8-1 Alarm information Bar
High-priority alarms need to be handled immediately.
Alarm information area
prompts
5 alarm sounds, 2 of which
are rapid, and the alarm
period is 9 seconds.
The red-bottom cycle
is shown.
3 tones, alarm period is 6
seconds.
The yellow-bottom
circulation is shown.
2 tones, with an alarm
period of 27 seconds.
True color and low
cycle display.
Attention
When muted, the alarm bell changes to and the alarm sound disappears. After
2 minutes, the alarm bell returns to its original state; if the conditions for the alarm
are not processed, the alarm sound will continue to sound..
8.2. Alarm Information Table
Attention Protect animal patients during use; repair the fault after use.
Attention This table does not include the operating instructions.
The cable not properly
connected
Power power outage
blown fuse
Check whether the cable
connection correct
Check if the power
supply interrupted
Check if the fuse
broken
If the fuse is
broken,
replace
immediately.