9.8.7. Alarm Performance
Alarm parameters
Setting range
Resolution
Upper pressure limit
0~ 60 cmH 2O
1 cmH 2O
Lower pressure limit
0~ 60 cmH 2O
1 cmH 2O
Upper limit of tidal
volume
0 ~ 1500mL
1mL
Lower limit of tidal
volume
0 ~ 1500mL
1mL
Upper limit of ventilation
0 ~ 40 L /min
.1 0L /min
Lower limit of ventilation
0 ~ 40 L /min
.1 0L /min
Upper limit of oxygen
concentration
15 ~ 100%
1%
Lower oxygen
concentration
15 ~ 100%
1%
PEEP upper limit
0~ 25 cmH 2O
1 cmH 2O
PEEP lower limit
0~ 20 cmH 2O
1 cmH 2O
Upper frequency limit
1 ~ 150bpm
1 bpm
Lower frequency limit
1 ~ 150bpm
1 bpm
Absorber heating
32-42
1
Apena time
3 ~ 20s
1s
Drive pneumatic
pressure is low
230±20 KPa
The alarm volume
2~ 10
1
Attention The lower limit of the above parameters cannot be set higher than the upper limit,
and the upper limit cannot be set lower than the lower limit.