TA80V User Manual
9.8.6. Monitoring Performance
Exhaled tidal
volume TV _ Exp
± 30 mL (less than 200 mL); ± 15%
(other)
Inhaled tidal
volume, TV _ Insp
± 30 mL (less than 200 mL); ± 15%
(other)
Display range is
from 0 to 99 L / min
The use range is
from 0 to 25 L / min
± 1 L/min (less than 5 L/min); ± 15%
(other)
Spontaneous
ventilation
MV_Spont
Display range is
from 0 to 99 L / min
The use range is
from 0 to 25 L / min
± 1 L/min (less than 5 L/min); ± 15%
(other)
2 bpm (below 20 bpm); ± 10% (other)
Spontaneous
frequency
Freq_Spont
2 bpm (below 20 bpm); ± 10% (other)
3 cmH
2
O (below 20 cmH
2
O); 15%
(other)
3 cmH
2
O (below 20 cmH
2
O); 15%
(other)
3 cmH
2
O (below 20 cmH
2
O); 15%
(other)
Inhaled oxygen
concentration of
FiO
2
2 cmH
2
O (below 10 cmH
2
O); 20%
(other)
Inspiratory to
expiratory I:E
Drive gas pressure
Press_Drive
5 KPa (below 50 KPa); 10% (other)
Pressure monitoring range: 0-80 cmH
2
O; X-axis: 0 to
10 seconds
Y-axis range: -120 ~ 100 L/min; X-axis range: 0 to 10
seconds.
Y-axis range: 0 to 800 mL; X-axis range: 0 to 10
seconds.