TA80V User Manual
50
Touch the Start and a
progress bar displays the
calibration.
Step 5
After the calibration
finished, a message
indicating whether the
verification is successful
or not will be displayed;
touch the button to return
to complete the pressure
sensor calibration.
7.3.3. O
2
Sensor Replacement
Warning
Please do not perform the calibration process while the system is connected to the animal.
1. When calibrating the O2 sensor, the ambient pressure must be the same as that used for
oxygen delivery monitoring in the animal circuit.
2. If the pressure during operation is different from the calibrated pressure, the monitoring
accuracy of the readings may exceed the specified range.
7.3.4. O
2
Sensor Correction
Warning
Do not perform the correction process when the system is connected to the animal patient.
check O
2
The ambient pressure must be the same as that used for monitoring of oxygen delivery in
the animal patient.
If the pressure during the operation is different from the formal pressure, the monitoring accuracy of
the reading may be beyond the specified range.
7.3.4.1. 21% Oxygen Sensor Calibrate