TA80V User Manual
6.4. CO2 Absorber
Inhalation valve and exhalation valve
Rotate the valve cover counterclockwise to loosen it. Use a water-soluble disinfectant to soak sterile
gauze to scrub the valve cover, valve cover sheet and valve opening. After drying the water, restore
the cover sheet, and tighten the valve cover clockwise. . After installation, the air tightness and
movement of the flap cover should be checked according to the debugging procedure. Be careful
when cleaning, and do not break or crush the valve cover and cover sheet.
CO2 absorber module
The disinfection method can be according to the conditions of the hospital, using fumigation (the
temperature is preferably not more than 50 ℃), disinfectant soaking, etc.; if the disinfectant is soaked,
after the disinfection is completed, the disinfected parts need to be blown dry with high-pressure air or
oxygen.
6.5. Bellows
This section is about the disassembly, assembly, cleaning and disinfection of the integrated
bellows.Understand this section thoroughly before starting disassembly, assembly, cleaning and
disinfection. Otherwise, the equipment will not work normally, and animal patients may be
endangered.
Warning
Only the material of the bellows folding bag is latex.
6.5.1. Remove the Bellow Components
The following are the steps to remove the bellows integration (the assembly steps are the reverse):
Release the screw on the bellows integration bracket and remove the bellows integration.
Rotate counterclockwise and remove the bellows cover