TA80V User Manual
37
6.1. Cleaning and Disinfection Before the First Use
Whole unit
Moist soft cloth with a water-soluble disinfectant used to clean
the ventilator panel and surface
Can be sterilized by ultraviolet rays, and peracetic acid and
formaldehyde fumigation are prohibited.
Breathing circuit and parts
See 6.2
Absorber
Cleaning and disinfection (see 6.4 for disinfection methods)
Bellow
Cleaning and disinfection (see 6.5.4 for disinfection methods)
6.2. Parts of the Breathing Circuit
Threaded tube , mask,
Y connector, L connector, air bag
This is a one-time design without disinfection. Waste should be
recycled.
These consumables should be replaced with medical grade
non-toxic, odorless, and the same size products
Reusable threaded tube, airbag
Soak with disinfectant
T connector
Soak with disinfectant
Flow sampling probe and flow
sampling tube
Every time a pet patient is replaced, it should be rinsed with
soapy water first, and then placed in a fumigation box for
disinfection after the water is dried.
6.3. Soda Lime Canister
Please refer to section 5.2.2 of this manual
Cleaning mechanical with detergent or washing disinfectant .
Put the absorption tank in detergent or washing disinfectant and clean it with decontamination
procedure
Put the absorption tank in the heating chamber, the maximum temperature is 80°C (176°F) or room
temperature
If the detergent or washing disinfectant cannot achieve the purpose of disinfecting the equipment, our
company recommends a higher level of disinfection
Manual cleaning
Rinse the absorption tank in running water
Completely immerse the absorption tank in the water tank filled with water and detergent for about 3
minutes. The water temperature should be approximately 40°C (104°F)
Rinse the absorption tank in running water
Our company recommends: After manual cleaning, a higher level of disinfection must be carried out
The third step: advanced disinfection
Before advanced disinfection, the absorption tank must be cleaned
The absorption tank can be placed under high temperature and high pressure vapor conditions. The
highest recommended temperature is 134°C (273°F)
After drying, pour the soda lime into the absorption tank, and then lock the handle tightly. Wipe off the
soda lime powder