TA80V User Manual
Touch the Start and a
progress bar displays the
calibration.
Step 5
After the calibration is
completed, a message
indicating whether the
calibration was successful
or not will be displayed;
touch the button to return
to complete the pure
oxygen sensor calibration.
If the calibration is
unsuccessful, you can
replace the oxygen
sensor, recalibrate the
21% oxygen, and then
calibrate the 100% oxygen
sensor.
7.3.5. Flow Rate Sensor Correction
The TA80V flow sensor uses differential pressure flow rate measurement. It can use the air
resistance on the loop or the differential pressure air resistance purchased at the animal end for
measurement. Different air resistances can be selected through settings. The settings are as follows.
Step 1
First switch to "Standby" ,
touch the "System "to view
the menu.