TA80V User Manual
Step 2
Step 3
Turn the knob clockwise to
ON; counterclockwise to
OFF. Select ON when
using outsourced air
resistance, and select
OFF when using the air
resistance provided by the
absorber.
Step 4
Press the knob to confirm
the selection.
The flow sensor has an inspiratory proportional coefficient and an expiratory proportional coefficient,
which are used to calibrate the inhaled and exhaled tidal volumes. These two coefficients have been
set before leaving the factory. If the circuit or air resistance is replaced, if the tidal volume error is
relatively large, it is necessary to Reset these two coefficients. The setting method is as follows.
Step 1
First switch to "Standby" ,
touch the "System " to
view the menu
Step 2
Step 3