TA80V User Manual
9.4. System Technical Specifications
9.4.1. Gas Supply
Gas supply
compression
release valve:
Oxygen and O2 connections not interchangeable.
Attention
All systems gas supply must be of medical grade.
When the anesthesia system stops delivering gas, the pipeline supply pressure is 280-600 kPa.
9.4.2. Flow Rate
Gas supply
Attention
The linkage ensures that the O
2
The output concentration is not less than 25%.
Accuracy: Under the conditions of 20℃ and 101.3kPa, for the flow rate between ±10% of full scale or
300mL/min (whichever is the larger value) to full scale, the accuracy is within ±10% of the indicated
value. Accuracy below 10% of full scale or 300mL/min (whichever is greater) is Level 4. Accuracy will
vary depending on breathing circuit pressure, barometric pressure or temperature. Under certain
conditions, these changes can exceed the allowable error range.
O2 Flush supply: 35-75 L/min
O2 supply failure alarm and stop
Attention O2 supply failure alarm takes precedence over AIR cutoff.