Remove the seal ring
6.5.2. Bellow Integration Functional Test
Warning Do not block the interface with small objects to prevent them from slipping into the
breathing circuit.
Warning
This functional test should be performed before use.
This test ensures that all components are assembled correctly. This test is not a substitute for system
testing. If the bellow integration meets the test requirements, it can be installed; otherwise, it must be
disassembled again, check whether any parts are damaged, and replace them in time before
assembly and testing.
Before installation, hand-held bellows integration, vertical upwards, blocks the drive gas interface
Inverted bellows integrated. The rate of descent of the top of the folded sac is not greater than
100mL/min. If the limit is exceeded, possible reasons include: the drive gas interface is not tightly
blocked, the folding bag or sealing ring is incorrectly installed, or other components are damaged.
Open the drive gas interface, fold the capsule fully deployed, and then block the breathing interface.