TA80V User Manual
43
1. Bellow cover
5. Lock spring
2. Folded bag
6. Seal ring
3. The tray
7. The base
4. Overflow valve
6.5.4. Cleaning and Disinfection
Clean and disinfect using methods recommended by the washer and sanitizer manufacturers.
Cleaning
1. Remove the bellows assembly.
Warning Do not separate the relief valve diaphragm and valve seat
2. Prevent damage to components. Wash gently and in hot water with a mild, enzyme-free
detergent recommended for rubber and plastics.
Attention Do not soak for more than 15 minutes to prevent swelling or aging.
3. Rinse with clean hot water and dry
Attention When drying the folded bag, it should be hung and fully unfolded. Otherwise,
adhesive capsulitis may result
4. After completely drying, check the parts for damage, then proceed to assembly and functional
testing
5. Connect the bellows integration, ventilator and respiratory syste.
6. Perform pre-use inspection of the system
Disinfection
In general, cleaning and disinfection should be carried out simultaneously. The following is only a
brief description of common methods that can be used for integrated disinfection of the bellows
integration.
Disinfection of General Pet Patients After Use
After animal patients use it, they should first wash the inside and outside with soapy water, rinse and
dry it repeatedly with clean water. Then, after soaking the plastic and rubber utensils in 70-80%
ethanol for half an hour, take them out with a sterile transfer forceps, store them in a clean container,
and repeat the sterilization before the next use. Metal, glass and other parts can be sterilized by
high-pressure steam. The high-pressure steam cooker is used to increase the steam pressure, and
the temperature in it will also increase, so that the bacterial protein will solidify quickly and the
sterilization effect will be rapid and reliable. For example, under 1.05kg/cm2 and steam pressure, the
temperature can be increased to 121°C and maintained for 15-20 minutes to kill all bacteria and most
of the spores.