TA80V User Manual
53
First switch to "Standby" ,
touch the "System " to
view the menu.
Step 2
Touch the "System
calibrate" to view the
menu.
Step 3
Touch the "Oxygen
Sensor" button. Note the
prompt message to place
the oxygen sensor in the
pure oxygen circuit for at
least 2 minutes.
Step 4

Table of Contents