TA80V User Manual
Step 4
Touch the Start and a
progress bar displays the
calibration.
Step 5
After the calibration is
completed, a message
indicating whether the
calibration is successful or
not will be displayed;
touch the button to return
to complete the calibration
of the oxygen sensor in
air. If the calibration is
unsuccessful, you can
replace the oxygen sensor
and calibrate again
7.3.4.2. 100% O2 Sensor Calibration
This process can take up to three minutes. Before you select pure oxygen concentration calibration,
you must complete air oxygen concentration calibration.
Warning Please do not perform the calibration process while the system is connected to the
animal
Before 100% oxygen concentration calibration, 21% O2 calibration must be done first.
Step 1