Warning Do not perform the correction process when the system is connected to the animal
patient.
The process takes a minimum of two minutes.
Before the pure oxygen and oxygen concentration is checked, the air and oxygen concentration
check must be done first.
Step 1
First switch to "Standby" ,
touch the "System " to
view the menu.
Step 2
Touch the "System
calibrate" to view the
menu.
Step 3
Touch the "21% O2
sensor calibrate". Note the
prompt to place the
oxygen sensor in air for at
least 2 minutes.