TA80V User Manual
60
High airway
pressure
High
Airway pressure value is
higher the upper limit set
value.
Tidal volume set at too
large volume.
The patient had a blocked
airway.
The expiration gas valve is
blocked.
Reset upper airway
pressure limit.
Check the exhalation
passage and deal with
blockages.
Check tidal volume
setting.
Check the patient's
airway and treat
obstructions.
----
Low airway
pressure
High
No driven gas.
Sampling tube is falling off
or blocked.
The breathing rate too low.
Reset the lower airway
pressure limit;
Check the sampling
tube.
----
No tidal volume
High
The tidal volume
monitored for 20 seconds
is 0.
Check animal patients.
Check the pipeline
connections.
----
High tidal
volume
High
The tidal volume is higher
than the set upper limit.
Check whether the
animal is breathing
spontaneously.
Check ventilator and
alarm settings
----
High minute
ventilation
High
The minute ventilation is
higher the upper limit.
Check whether the
animal is breathing
spontaneously.
Check ventilator and
alarm settings
----
Low minute
ventilation
High
Minute ventilation is lower
the lower limit.
Leakage occur.
Reset the lower minute
ventilation limit.
Check the animal end.
Check pipe connections.
----
Apena
High
No airflow passes through
within the set time
Check the animals.
Switch to manual mode
and make ventilation by
hand.
Check if the connection
is broken.
----