4
Operator's Manual English 37
4. Operation
4.1. Overview
This chapter explains how to use the INVOS™ patient monitor (the “monitoring system”) to
view and collect patient regional oxygen saturation (rSO
2
) data.
This chapter assumes that the monitoring system has been installed at the location of use
and tested by the institution. See Installation, Chapter 3, Installation, page 23 for full
installation instructions.
•
Safety Reminders, page 37
•
Quick Start, page 39
•
Operational Defaults, page 40
•
Menu Structure, page 41
•
Monitoring System Memory, page 42
•
Set Up for Patient Monitoring, page 42
•
Optional Set-Up Tasks, page 46
•
Baselines, page 52
•
Trend View Management, page 55
•
Alarm Management, page 60
•
Event Marks, page 66
•
Area Under the Curve (AUC), page 70
•
Finish Monitoring, page 72
•
Case Histories, page 73
4.2. Safety Reminders
Warning: The monitoring system should not be used as the sole basis for diagnosis or
therapy. It is intended only as an adjunct in patient assessment. It must be used in
conjunction with clinical signs and symptoms.
Warning: Inspect the monitoring system and all accessories before use to ensure there are
no signs of physical damage or improper function. Do not use if damaged.
Warning: Explosion hazard — Do not use the monitoring system in the presence of
flammable anesthetics.
Warning: Always disconnect and remove the monitoring system and sensors during
magnetic resonance imaging (MRI) scanning. Attempting to use the monitoring system
during an MRI procedure could cause burns or adversely affect the MRI image or the
monitoring system's performance.
Warning: As with all medical equipment, carefully route patient cabling to reduce the
possibility of patient entanglement or strangulation.