EasyManua.ls Logo

Medtronic INVOS - Operational Defaults; Table 6. Operational Defaults

Medtronic INVOS
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
40 Operator's Manual English
Table 5. Quick Start Steps (continued)
17 Respond to alarms, as necessary:
Alarm indicators
Alarm message lookup
Silence or pause alarms
Dismiss alarms
See:
Alarm Indicators, page 60
Alarm Messages, page 99
Silencing or Pausing Alarms, page 64
Dismissing an Alarm, page 65
18 Mark events, as desired See Event Marks, page 66
19 Finish monitoring See Finish Monitoring, page 72
4.4. Operational Defaults
The following table lists default settings for the monitoring system. Some of these settings
can be changed by the user, while others require authorized access, as indicated.
Table 6. Operational Defaults
Item Available settings Default
Upper alarm limits OFF, ON OFF
See Setting Alarm Limits, page 62
Lower alarm limits MANUAL, AUTO MANUAL
See Setting Alarm Limits, page 62
Upper rSO
2
limit 20 - 95 If ON, default is 90
See Setting Alarm Limits, page 62
Lower rSO
2
limit MANUAL: 15 - 90
AUTO: 5 - 30
MANUAL: 40
AUTO: 20
See Setting Alarm Limits, page 62
Silence alarms at startup YES, NO NO - Alarm audio is on by default, but alarm audio may
be silenced or paused by the user. See Silencing or
Pausing Alarms, page 64.
This setting may be changed by an authorized techni‐
cian.
Alarm audio off/ pause
duration
INDEFINITE, 2 MINS INDEFINITE - When silenced, alarms remain silent until
the user unsilences them. See Alarm Silence Duration,
page 65.
This setting may be changed by an authorized techni‐
cian.
Alarm reminder signal ON, OFF OFF - No reminder when alarms are silenced
This setting may be changed by an authorized techni‐
cian.
Line frequency 50 Hz, 60 Hz 60 Hz
This setting may be changed by an authorized techni‐
cian.
Alarm volume 1 - 10 5
See Changing the Alarm Volume, page 63
Screen brightness 1 - 10 7
See Changing the Screen Brightness, page 51
Time format 24 hr, 12 hr 12 hr
This setting may be changed by an authorized techni‐
cian.

Table of Contents

Other manuals for Medtronic INVOS

Related product manuals