7
Operator's Manual English 95
7. Product Maintenance
7.1. Overview
This chapter describes the steps required to properly clean the INVOS™ patient monitor
(the “monitoring system”). It also provides information about periodic safety checks,
service, software and firmware upgrades, and component disposal.
Safety Reminders, page 95
Cleaning the Monitoring System, page 95
Maintenance Schedule, page 96
Service and Calibration, page 97
Service Life, page 97
Software and Firmware Updates, page 97
Recycling and Disposal, page 97
7.2. Safety Reminders
Warning: Inspect the monitoring system and all accessories before use to ensure there are
no signs of physical damage or improper function. Do not use if damaged.
Warning: Explosion hazard — When replacing the battery, do not use the incorrect type.
Use only the battery available from Medtronic. See Accessories/Parts List, page 107.
Warning: To ensure proper performance, avoid shock, and prevent device damage or
failure, do not expose the monitoring system to extreme moisture, such as direct exposure
to rain. Do not immerse in water, solvents, or cleaning solutions, since the monitoring
system and connectors are not waterproof.
Caution: Do not autoclave or gas sterilize any components of the monitoring system.
Caution: To prevent device damage or failure, do not expose the monitor to isopropyl
alcohol.
Caution: Dispose of the battery in accordance with local guidelines and regulations.
7.3. Cleaning the Monitoring System
Warning: To ensure proper performance, avoid shock, and prevent device damage or
failure, do not expose the monitoring system to extreme moisture, such as direct exposure
to rain. Do not immerse in water, solvents, or cleaning solutions, since the monitoring
system and connectors are not waterproof.
Caution: Do not autoclave or gas sterilize any components of the monitoring system.
Caution: To prevent device damage or failure, do not expose the monitor to isopropyl
alcohol.