Operator's Manual English 93
Disruption may be evidenced by cessation of operation or other incorrect functioning. If
this occurs, survey the site of use to determine the source of this disruption, then take the
appropriate actions to eliminate the source.
Turn equipment in the vicinity off and on to isolate the interfering equipment.
Reorient or relocate the interfering equipment.
Increase the separation between the interfering equipment and the monitoring
system.
Connect the monitoring system to an outlet on a different circuit from the other
device(s).
Have an authorized technician check the line frequency setting for the monitoring
system. The setting should match the AC power line input. Refer to the monitoring
system's service manual.
The monitoring system generates, uses, and can radiate radio frequency energy and, if not
installed and used in accordance with these instructions, may cause harmful interference
with other susceptible devices in the vicinity. Contact Technical Services for assistance.
See Obtaining Technical Assistance, page 10.