63
<Command>
*IDN?
FunctionIt queries the device basic data.
Description
Return data
MICROTEST,6632-30G,0,1.194<field1>,<field2>,<field3>,<field4>
<field1>manufacturer
<field2>model number
<field3>serial number or 0
<field4>firmware revision
MICROTEST,6632-30G,0,1.194
<Command> *OPC
FunctionIt sets Standard Event Status Register to 0 bit,when all operations are completed.
<Command> *OPC?
FunctionWhen all operations are completed, 1 will be returned.
<Command>
*OPT?
FunctionIt queries the installed in the system interface.
Description
If 6632 MAX test frequency is 1MHz , it returnsF01
If 6632 MAX test frequency is 3MHz, it returns F03
If 6632 MAX test frequency is 5MHz, it returns F05
If 6632 MAX test frequency is 10MHz, it returns F10
If 6632 MAX test frequency is 15MHz, it returns F15
If 6632 MAX test frequency is 20MHz, it returns F20
If 6632 MAX test frequency is 30MHz, it returns F30
<Command>
*RST
FunctionIt aborts all pending operations, and sets 6632 to its initial setups.
Description
Initial Setups
Parameters:Ls,Q,Z,θdeg
FREQuency:1kHz
LEVEL:1Vac
SPEED:MED.
TRIGger:REPEAT
<Command>
*SRE <NR1>
*SRE?
FunctionIt sets or queries the status of Service Request Enable Register.
Description
Set parameter 0~255
Set syntax
*SRE 1
Query syntax
*SRE?
Return data 0~255 (Format is in <NR1>)
<Command>
*STB?
FunctionIt queries the status of Service Request Status Register
Description
Return data<NR1>

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Microtest 6632 Series and is the answer not in the manual?

Microtest 6632 Series Specifications

General IconGeneral
BrandMicrotest
Model6632 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals