83
<Command>
:SWEep:FUNCtion
{LS|LP|CS|CP|Q|D|RS|RP|Z|DEG|RAD|R|X|Y|G|B,OFF|LS|LP|CS|CP|Q|D|RS|RP|Z
|DEG|RAD|R|X|Y|G|B }
:SWEep:FUNCtion?
FunctionIt sets or queries the FUNC.
Description
Set parameter
The comma is placed between the two parameters. The first parameter can not
set to OFF. There is no RDC(DC Resistance),
E(Relative Permittivty) and U(Relative
permeability) in the sweep mode.
Set syntax
:SWEep:FUNCtion Z,DEG
Querysyntax
:SWEep:FUNCtion?
Return
data Z,DEG(Format is in <disc>)
<Command>
:SWEep:ANALysis:MODel {A|B|C|D|E|F|G|OFF}
:SWEep:ANALysis:MODel?
FunctionIt sets or queries the ANALYSIS MODEL.
Description
Set syntax
:SWEep:ANALysis:MODel B
Query syntax
:SWEep:ANALysis:MODel?
Return data B(Format is in <disc>)
<Command>
:SWEep:DELay <delay timeNR3/disc>
:SWEep:DELay?
FunctionIt sets or queries the SWEEP DELAY.
Description
Set parameterThe value of delay time is {0.0~5|MAXimum|MINimum}.
Set syntax
:SWEep:DELay 0.5
:SWEep:DELay 500M
:SWEep:DELay 500MS
:SWEep:DELay 5E-3
:SWEep:DELay MAXimum
:SWEep:DELay MINimum
Query syntax
:SWEep:DELay?
Return
data 0.500 (Format is in <NR2>)
<Command>
:SWEep:OIMPedance {100|25}
:SWEep:OIMPedance?
FunctionIt sets or queries the OUTPUT IMPEDANCE.
Description
Query syntax
:SWEep:OIMPedance?
Return data 100|25 (Format is in <NR1>)
<Command>
:SWEep:KEEP {OFF|ON|0|1}
:SWEep:KEEP?
FunctionIt sets or queries the KEEP PREVIOUS TRACE.
Description
Set syntax
:SWEep:KEEP ON
:SWEep:KEEP 0
Querysyntax
:SWEep:KEEP?
Return
dataON | OFF (Format is in <disc>)