EasyManua.ls Logo

Micsig TO1102 - Page 21

Micsig TO1102
169 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18
Figure 2-13 Probe Connection
3) Open the channel (if the channel is closed).
4) Adjust the oscilloscope channel attenuation coefficient to match the probe attenuation ratio.
5) Tap button or manually adjust the waveform vertical sensitivity and horizontal time base.
Observe the shape of the waveform, see Figure 2-14.
Figure 2-14 Probe Compensation
If the waveform on the screen is shown as “under-compensation” or “over-compensation”, please adjust the
trimmer capacitor until the waveform shown on the screen as “correct-compensation”. The probe adjustment is
shown in Figure 2-15.

Table of Contents

Related product manuals