Further settings
66
A0 value display
The A0 value is a measure for the
deactivation of micro-organisms in
disinfection processes using moist heat.
This value is used in thermal disinfection
processes in machine-based reprocessing
to determine the amount of moist heat
needed.
The A0 value is displayed at the end of the
programme.
08:15
Off
On
Display A0 value
- On
A0 value measuring is switched on and
the value is displayed at the end of the
programme
- Off
A0 value measuring is switched off
Select an option.
Dispensing systems
The following menu allows you to activate,
fill and rinse dispensing systems, to change
their names if required and to set the
dispensing concentration for all
programmes.
Depending on the model, various
dispensing systems are available for
dispensing process chemicals.
Cleaning agent and rinsing agent can be
dispensed either via a corresponding door
dispenser (cleaning agent, rinsing
agent) or via suction lances for liquid media
from an external canister. It is not possible
to activate both systems at the same time.
Selecting a dispensing system
The menu is saved under the following input
path.
Extended settings
Dispensing systems
Cleaning agent door
Clean. agent canister
Rinse aid door
Rinse aid canister
08:15Dispensing systems
Cleaning
agent door
Rinse aid
door
Rinse aid
canister
Clean. agent
canister
Only the available dispensing systems are
displayed.
Select a dispensing system.