Milli-Q® Direct 8/16 System 19
Specifications and requirements, Continued
The Feed water requirements are listed here.
The Environmental requirements are listed here.
Ambient operating temperature
Ambient storage temperature
The System is intended for indoor
use only.
Relative humidity during storage
and operation
Maximum relative humidity 80% for
temperatures up to 31°C decreasing
linearly to 50% relative humidity at
The noise level is < 50 dB at a distance of 1 metre.
The minimum consumables required for installation are listed here.
Note that these items are not shipped with the System and must be ordered separately:
• Progard® Cartridge,
• Q-Pak® Pack, and
•
location
The Reservoir must be located relative to the Water System:
• 0 ≤ y ≤ 2 metres, where y = vertical distance, and
•
≤
≤
3 metres, where x = horizontal distance.