Milli-Q® Direct 8/16 System 21
Alarms generated during installation
During the installation of a Milli-Q® System, certain Alarm messages are generated.
• This occurs because:
o the Reservoir is empty,
o there is air in the tubings and in the Progard® Cartridge,
o the Progard® Cartridge is not installed, and
o the Q-Pak® Pack is not installed.
• These alarms are explained here. For more information about Alarm messages, see the
message
This alarm occurs because the Reservoir is empty during most of the installation.
• This alarm goes away when the Reservoir is partially full.
•
To cancel the text display of this alarm message, follow the instructions on the LCD.
CARTRIDGE OUT
message
This alarm occurs because the Progard® Cartridge is not installed.
• This alarm goes away when the Progard® Cartridge is detected by the Milli-Q®
System.
•
To cancel the text display of this alarm message, follow the instructions on the LCD.
OUT message
This alarm occurs because the Q-Pak® Pack is not installed.
• This alarm goes away when the Q-Pak® Pack is detected by the System.
•
To cancel the text display of this alarm message, follow the instructions on the LCD.
SP, REPLACE
Q-PAK message
This alarm occurs because the Q-Pak® Pack is not fully rinsed out or there is air in the
tubing near the resistivity sensor.
• This alarm goes away when a few litres of water are dispensed from the
POD Unit.
• To cancel the text display of this alarm message, follow the instructions on the LCD.
WATER
PRESSURE
message
This alarm occurs because there is air in the tubings and in the new Progard®
cartridge.
• When the air is gone and replaced with water, this alarm does not occur anymore
during installation.
•
To cancel the text display of this alarm message, follow the instructions on the LCD.
Continued on next page