EasyManua.ls Logo

Milli-Q Direct 8 - Alarms Generated During Installation

Milli-Q Direct 8
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Milli-Q® Direct 8/16 System 21
Alarms generated during installation
Overview
•
During the installation of a Milli-Q® System, certain Alarm messages are generated.
• This occurs because:
o the Reservoir is empty,
o there is air in the tubings and in the Progard® Cartridge,
o the Progard® Cartridge is not installed, and
o the Q-Pak® Pack is not installed.
• These alarms are explained here. For more information about Alarm messages, see the
chapter titled ‘Alarms’.
TANK EMPTY
message
•
This alarm occurs because the Reservoir is empty during most of the installation.
• This alarm goes away when the Reservoir is partially full.
•
To cancel the text display of this alarm message, follow the instructions on the LCD.
PROGARD
CARTRIDGE OUT
message
•
This alarm occurs because the Progard® Cartridge is not installed.
• This alarm goes away when the Progard® Cartridge is detected by the Milli-Q®
System.
•
To cancel the text display of this alarm message, follow the instructions on the LCD.
Q-PAK PACK
OUT message
•
This alarm occurs because the Q-Pak® Pack is not installed.
• This alarm goes away when the Q-Pak® Pack is detected by the System.
•
To cancel the text display of this alarm message, follow the instructions on the LCD.
MILLI-Q RES <
SP, REPLACE
Q-PAK message
•
This alarm occurs because the Q-Pak® Pack is not fully rinsed out or there is air in the
tubing near the resistivity sensor.
• This alarm goes away when a few litres of water are dispensed from the
POD Unit.
• To cancel the text display of this alarm message, follow the instructions on the LCD.
LOW FEED
WATER
PRESSURE
message
•
This alarm occurs because there is air in the tubings and in the new Progard®
cartridge.
• When the air is gone and replaced with water, this alarm does not occur anymore
during installation.
•
To cancel the text display of this alarm message, follow the instructions on the LCD.