Milli-Q® Direct 8/16 System 37
Registering UV Lamp timers
Introduction
The timer used for each UV Lamp must be reset when the System is installed.
If this is not done, then the message indicating that a Lamp replacement is needed is
shown too early.
The UV Lamp timers need to be reset for:
the UV 185 nm Lamp, and
the UV 254 nm Lamp.
NOTE:
Before doing this, make sure that the date and time have been checked for accuracy.
Procedure
This procedure shows how to reset the timer used for the UV 185 nm Lamp.
Step
Action
Diagram
1
Place the System in STANDBY
Mode.
2
Select Menu.
Press .
3
Select Maintenance.
Press .
4
Select Install UV 185 nm Lamp.
Press .
Continued on next page

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Milli-Q Direct 8 and is the answer not in the manual?

Milli-Q Direct 8 Specifications

General IconGeneral
Resistivity18.2 MΩ·cm at 25 °C
Total Organic Carbon (TOC)< 5 ppb
Pyrogens (Endotoxins)< 0.001 EU/mL
RNases< 1 pg/mL
DNases< 5 pg/mL
DisplayTouchscreen
Weight20 kg
Power Supply100-240 V, 50/60 Hz
Type of Water ProducedPure and ultrapure water
Flow RateUp to 2 L/min
Particulates<1 particle/mL (>0.2 µm)
Particles > 0.22 μm< 1 particulate/mL
Particle Size0.22 µm
ConnectivityUSB
Output Water TypeUltrapure
Dimensions504 mm x 332 mm x 467 mm (19.8 in x 13.1 in x 18.4 in)

Related product manuals