EasyManuals Logo

Milli-Q Direct 8 User Manual

Milli-Q Direct 8
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
Milli-Q® Direct 8/16 System 49
Description of Standby Menu
Maintenance
The Maintenance Menu is described below.
Diagram 1
Diagram 2
Item
Description
Install Pretreatment
Used to reset Alert message REPLACE EXTERNAL
PRE-TREATMENT.
Clean Strainer
Used to reset Alert message EXAMINE INLET
STRAINER.
Install Progard®
Used to see general information about the Progard®
Cartridge exchange.
Install new RO
Used to start a flush and rinse of a new RO
Cartridge.
Install UV 254 Lamp
Used to reset Alert message REPLACE 254 NM
LAMP.
Install UV 185 Lamp
Used to reset Alert message REPLACE 185 NM
LAMP.
Install Q-Pak®
Used to see general information about the Q-Pak®
pack exchange.
Install POD Pak
Used to reset Alert message REPLACE POD PAK.
Install ASM UV
Used to reset Alert message REPLACE ASM UV
LAMP
Reset RO CL2 Cleaning
Used to reset Alert message PERFORM RO CL2
CLEANING at installation.
Continued on next page

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Milli-Q Direct 8 and is the answer not in the manual?

Milli-Q Direct 8 Specifications

General IconGeneral
Resistivity18.2 MΩ·cm at 25 °C
Total Organic Carbon (TOC)< 5 ppb
Pyrogens (Endotoxins)< 0.001 EU/mL
RNases< 1 pg/mL
DNases< 5 pg/mL
DisplayTouchscreen
Weight20 kg
Power Supply100-240 V, 50/60 Hz
Type of Water ProducedPure and ultrapure water
Flow RateUp to 2 L/min
Particulates<1 particle/mL (>0.2 µm)
Particles > 0.22 μm< 1 particulate/mL
Particle Size0.22 µm
ConnectivityUSB
Output Water TypeUltrapure
Dimensions504 mm x 332 mm x 467 mm (19.8 in x 13.1 in x 18.4 in)