EasyManua.ls Logo

Milli-Q Direct 8 - Cleaning the RO Cartridge(S)

Milli-Q Direct 8
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
90 Milli-Q® Direct 8/16 System
Cleaning the RO Cartridge(s)
When
Cleaning the RO cartridge may be required when feed water contains excess of some
chemicals.
Perform cleaning if required, after recommendation from your Millipore SAS Service
Representative.
Safety
Wear Eye Safety Glasses and Laboratory Gloves and other appropriate safety equipment
when cleaning the RO Cartridge(s).
Cleaning
Follow the steps below to clean the RO Cartridge(s).
Step
Action
Diagram
1
Open the sanitisation port as
described in the previous
section.
2
Place a cleaning agent pouch
(ROCleanA or ROClean B)
into the Sanitisation Port.
3
Put the cap back on and hand-
tighten it.
NOTE:
The chemical in the pouch will
dissolve during the pH Cleaning
sequence. Remove the empty
pouch the next time the
Santisation Port cap is removed.
4
Go to STANDBY Mode.
Continued on next page
!

Related product manuals