110 Milli-Q® Direct 8/16 System
Summary of Alert messages
Alert messages
LCD message
What it means
CALIBRATION VISIT OVERDUE XX
DAYS
The System has determined that a
Calibration Visit is overdue.
Contact Millipore SAS.
CHECK ASM UV LAMP
The ASM UV Lamp is not turning on.
Contact Millipore SAS.
CHECK UV 185 NM LAMP
The UV 185 nm Lamp is not turning on.
Contact Millipore SAS.
CHECK UV 254 NM LAMP
The UV 254 nm Lamp is not turning on.
Contact Millipore SAS.
EXAMINE INLET STRAINER
The System has determined that it is
time to clean the Inlet Strainer.
Clean the Inlet Strainer and reset the
message.
LOW RO PUMP PRESSURE
The Water System has determined that
the RO Pump Pressure is below
specification.
Contact Millipore SAS.
NEXT CALIBRATION VISIT IN XX
DAYS
The System is prompting you that a
Calibration Visit should be scheduled.
Contact Millipore SAS.
NEXT QUALIFICATION VISIT IN XX
DAYS
The System is prompting you that a
Qualification Visit should be scheduled.
Contact Millipore SAS.
NEXT SERVICE VISIT IN XX DAYS
The System is prompting you that a
Service Visit should be scheduled.
Contact Millipore SAS.
NO RESPONSE FROM DHCP
SERVER
Contact your network administrator.
Restart the System.
PERFORM RO CL2 CLEANING
The Water System has determined that
it is time to perform a RO Cl2 cleaning.
Start a RO Cl2 cleaning. The timer is
reset automatically.
Continued on next page