Milli-Q® Direct 8/16 System 85
TOC Curve Check
When
A TOC Curve Check should be done when the Q-Pak® pack is replaced or when TOC
values are fluctuating to ensure that the TOC values displayed are valid.
NOTE:
The TOC Curve Check can be started manually following the replacement and flushing
of the Q-Pak® pack. The System automatically performs a TOC Curve Check once per
day. So, if the TOC Curve Check is not started manually following the consumables
replacement, it will be done automatically within the next 24 hours or sooner.
Procedure
Follow the steps below to perform a TOC Curve Check.
Step
Action
Diagram
1
Go to READY Mode.
2
Select Menu.
Press .
3
Select TOC Curve Check.
Press .
Continued on next page

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Milli-Q Direct 8 and is the answer not in the manual?

Milli-Q Direct 8 Specifications

General IconGeneral
Resistivity18.2 MΩ·cm at 25 °C
Total Organic Carbon (TOC)< 5 ppb
Pyrogens (Endotoxins)< 0.001 EU/mL
RNases< 1 pg/mL
DNases< 5 pg/mL
DisplayTouchscreen
Weight20 kg
Power Supply100-240 V, 50/60 Hz
Type of Water ProducedPure and ultrapure water
Flow RateUp to 2 L/min
Particulates<1 particle/mL (>0.2 µm)
Particles > 0.22 μm< 1 particulate/mL
Particle Size0.22 µm
ConnectivityUSB
Output Water TypeUltrapure
Dimensions504 mm x 332 mm x 467 mm (19.8 in x 13.1 in x 18.4 in)

Related product manuals