Milli-Q® Direct 8/16 System 101
Summary of Alarm messages
messages
This alarm stops the Milli-Q® portion
of the system.
• A safety feature of the Milli-Q® D
system has automatically stopped
dispensing water after 40 minutes to
avoid risks of overflow.
• Push the POD Unit Plunger all the
way down and release.
• This resets the dispenser timer and
makes the POD Unit available for
INCORRECT PROGARD CARTRIDGE
This alarm stops the RO portion of
the system.
• The System does not recognize the
type of Progard® Cartridge being
installed.
•
• This alarm stops the Milli-Q® portion
of the system.
• The System does not recognize the
type of Q-Pak® Pack being installed.
•
This alarm stops the RO portion of
the system.
• Check Feed water pressure and
rectify.
• Go to STANDBY Mode and go to
READY Mode to release any trapped
air in the Water System.
• Contact Millipore SAS if the problem
This alarm stops the RO portion of
the system.
• The Permeate conductivity is above
the set point.
•
• This alarm stops the Milli-Q® portion
of the system.
• The POD Unit was left in the open
position.
• Push the Plunger all the way down
Continued on next page