102 Milli-Q® Direct 8/16 System
Summary of Alarm messages, Suite
(suite)
This alarm stops the RO portion of
the system.
• The Progard® Cartridge is not
installed correctly or it has been
removed.
• Verify that the Progard® Cartridge is
installed correctly.
• Contact Millipore SAS if the problem
continues.
This alarm stops the Milli-Q® portion
of the system.
• The Q-Pak® Pack is not installed
correctly or it has been removed.
• Verify that the Q-Pak® Pack is
installed correctly.
• Contact Millipore SAS if the problem
• This alarm stops the Milli-Q® portion
of the system.
• The System has detected an empty
Reservoir.
• Refill the Reservoir.
• Verify that the Reservoir level sensor
is plugged into the System Cabinet.
This alarm stops the whole system to
prevent risks of flood.
• A Water Sensor (an accessory
connected to the System) has
detected water on the surface where
it is located. This may be caused by
a leak.
• Clean up the spilled water.
• Make sure the source of the leak is
fixed.
• Place the system in Standby Mode,
Ce sujet continue page suivante