Milli-Q® Direct 8/16 System 103
Summary of Alarm messages, Continued
Alarm messages
CLEANING CANCELLED
A cleaning mode was cancelled and
was not fully completed.
Go to STANDBY Mode and then go
to READY Mode.
The Milli- System will go into a
15 minute FLUSH Mode. The
system will then automatically fill
the Reservoir.
MILLI-Q INTER R > MAX
The Intermediate resistivity is out
of measurement range.
Contact Millipore SAS.
MILLI-Q INTER R < MIN
MILLI-Q INTER T < MIN
The Intermediate temperature is
out of measurement range.
Contact Millipore SAS.
MILLI-Q INTER T > MAX
MILLI-Q RES < SP, REPLACE Q-PAK
The Milli- Water resistivity is
below the set point.
Dispense water to eliminate any
trapped air in the System.
If the issue persists, replace the Q-
Pak® Pack.
MILLI-Q RES > MAX
The Milli- Water resistivity is out
of measurement range.
Contact Millipore SAS.
MILLI-Q T < MIN
The Milli- Water temperature is
out of measurement range.
Contact Millipore SAS.
MILLI-Q T > MAX
MILLI-Q TOC > SP
The TOC is above the set point.
Contact Millipore SAS.
PERMEATE C < MIN
The Permeate conductivity is out of
measurement range.
Contact Millipore SAS.
PERMEATE C > MAX
RO FEED C < MIN
The Feed water conductivity is out
of measurement range.
Contact Millipore SAS.
RO FEED C > MAX
RO FEED T < MIN
The Feed water temperature is out
of measurement range.
Contact Millipore SAS.
RO FEED T > MAX