Milli-Q® Direct 8/16 System 111
Summary of Alert messages, Continued
Alert messages
(continued)
LCD message
What it means
PERFORM TOC CURVE CHECK
The System has determined that a
TOC Curve Check should be
performed.
Perform a TOC Curve Check or wait
until one is done automatically.
If this message persists after 24 hours
of its appearance, then contact
Millipore SAS.
QUALIFICATION VISIT OVERDUE XX
DAYS
The System has determined that a
Qualification Visit is overdue.
Contact Millipore SAS.
REPLACE ASM UV LAMP
The Water System has determined
that the ASM UV Lamp should be
replaced.
Contact Millipore SAS.
REPLACE ASM UV LAMP IN XX
DAYS
The Water System has determined
that the ASM UV Lamp on the
Reservoir should be replaced in XX
days, where XX is 15, …, 1.
Contact Millipore SAS.
REPLACE EXTERNAL
PRETREATMENT
The Water System has determined
that the external pretreatment
(optional) should be replaced.
Consult the documentation supplied
with the external pretreatment for
more information.
REPLACE EXTERNAL
PRETREATMENT IN XX DAYS
The Water System has determined
that the external pretreatment
(optional) should be replaced in XX
days, where XX is 15, …, 1.
Consult the documentation supplied
with the external pretreatment for
more information.
REPLACE POD PAK
The System has determined that the
POD PAK needs replacement.
Replace the POD Pak and reset the
timer.
REPLACE POD PAK IN XX DAYS
The System has determined that the
POD PAK should be replaced in XX
days, where XX is 15, …, 1.
Replace the POD Pak and reset the
timer.
Continued on next page