A7â„¢ Service Manual 046-006272-00 3 - 11
Periodic Maintenance Functional Tests
The following screen is displayed if the safety valve control test is failed. You can select Service Access
and enter the required service password to access service mode. You can select Retry to perform
automatic circuit leak test again.
FIGURE 3-9 Machine Non-Functional Screen