Functional Tests Periodic Maintenance
3 - 12 046-006272-00 A7â„¢ Service Manual
The following screen is displayed if the automatic circuit leak test is failed and the Leak ≥200 mL/min
and ≤1000 mL. You can select Retry to perform the automatic circuit leak test again or accept the
result continue with automatic ventilation.
FIGURE 3-10 Automatic Circuit Leak and Compliance Test: Fail
The following screen is displayed if the automatic circuit leak test is failed and the leak >1000 mL. You
can select Manual Only to enter Standby mode. But mechanical ventilation is disabled. You can select
Retry to perform automatic circuit leak test again.
FIGURE 3-11 Automatic Circuit Leak and Compliance Test: Fail