Functional Tests Periodic Maintenance
3 - 30 046-006272-00 A7™ Service Manual
FIGURE 3-35 Calibration Device (VT Plus)
FIGURE 3-36 TSI Certifier 4070
2. Set up the calibration device as described below.
To set the Fluke VT-Plus:
a. Pressure Setting: Press the PRESSURE button on the front control panel of the calibration,
select Range and then set it to Low Press.