System Calibration Calibration
4 - 6 046-006272-00 A7â„¢ Service Manual
3. Set up the machine as per the instructions on the screen and select Next to open the menu
shown below.
FIGURE 4-2
4. Select Begin to calibrate flow sensors. During the calibration, you can select Cancel to cancel the
calibration.
FIGURE 4-3
5. The screen shown below is displayed if the ongoing calibration is canceled. Select Try Again to
do the calibration again. Select Done to exit the calibration screen.